false
OasisLMS
Login
Catalog
ASGE DDW Videos from Around the World | 2026
ENDOSCOPIC SALVAGE OF TYPE III STENT MISDEPOLYMENT ...
ENDOSCOPIC SALVAGE OF TYPE III STENT MISDEPOLYMENT DURING EDGE, USING NOTES TECHNIQUE
Back to course
[Please upgrade your browser to play this video content]
Video Transcription
Video Summary
During an EDGE procedure in a 62-year-old woman with prior Roux-en-Y gastric bypass and a common bile duct stone, a rare type 3 lumen-apposing metal stent misdeployment occurred: the distal flange entered the excluded stomach, but the proximal flange deployed into the peritoneal cavity. Using a NOTES-based salvage technique with an ultrathin endoscope, wires were re-established, spilled gastric contents were suctioned, and a second LAMS was placed successfully to create a gastroenteric conduit. The patient recovered well, later underwent ERCP with stone removal, and had no leak or major complications.
Asset Subtitle
M. Hayat, M.K. Hasan, N.D. Cosgrove, D. Jain, K. Kadhkodayan, D. Yang, M.A. Arain
Meta Tag
Concept
Lumen-Apposing Metal Stent Misdeployment
Concept
Type 3 Misdeployment
Concept
NOTES/NODES Technique
Concept
Endoscopic Ultrasound-Directed Gastroenterostomy
Concept
Excluded Stomach
Keywords
EDGE procedure
lumen-apposing metal stent
LAMS misdeployment
Roux-en-Y gastric bypass
NOTES salvage technique
Lumen-Apposing Metal Stent Misdeployment
Type 3 Misdeployment
NOTES/NODES Technique
Endoscopic Ultrasound-Directed Gastroenterostomy
Excluded Stomach
×
Please select your language
1
English